For the weighing system FES 17K-4:
For the weighing systems FES 33K-4 and FES 62K4-D:
Completing the adjustment:

19.1.2 Adjustment test with external weight

75
 Message <on FS> appears on the display and starts
flashing
 Adjustment starts
 Various messages are shown on the display one
after another: <on FS> → <push M>
 Press [MENU] key
 Message <on FS> appears on the display and starts
flashing
 Adjustment starts
 When <631 EX CAL> is displayed on the scale, the
adjustment is completed.
 Unload weighing plate
 Open <63 MAINTENANCE> (see chapter 19).
 Navigate to <632 EX SPAN TEST> using the [F1]
and [F2] keys.
 Press [F4] key to start external adjustment test
 The scale performs a zero setting
 Various messages are shown on the display one
after another: <buSY> → <on 0> → <on FS>
 With <on FS> the zero setting is completed.
 Place the adjustment weight centrally on the
weighing plate.
TFES_TFEJ-BA-e-2210