Weighing With A Tolerance Range; Stop Weight Value (Freeze) - KERN HTS Series Operating Instructions Manual

Electronic crane scale
Table of Contents

Advertisement

Returning to the peak weight measuring cycle
When the peak weight is displayed, press the C key. The display changes to the
current gross weight and the peak weight measuring cycle is ended.
Clearing the peak weight
1. When the peak weight is displayed, press the a keys. The display returns to the
actual gross weight.
2. Press the C key. The stored peak value is deleted and a new measuring cycle can
be started.
6.10

Weighing with a tolerance range

To ensure that your weighing value is within a defined tolerance range, you can
program separate upper and lower limits. Optical weighing help can support you when
using tolerance weighing. If optical weighing help is active, then the weight display will
flash when the weight is below the lower limit or above the upper limit.
Weighing with a tolerance range must be activated in the User menu, see chapter 7.4.
Defining limits
When "Weighing with a tolerance range" is selected as the weighing mode, then the
parameters F3.6.1 to F3.6.3 will be available in the user menu, see chapter 7.4.
1. Entering the upper limit, F3.6.1.
2. Entering the lower limit, F3.6.2
3. Activate/deactivate optical weighing help F3.6.3.
6.11

Stop weight value (freeze)

Once a stable weight value is detected, you can "freeze" or lock the weight value
displayed until it is cleared for the next weighing.
Press the buttons Mode and F to "freeze" or hold the current weight value.
It will be displayed as long as it isn't cleared.
To clear the "frozen" or locked weight, press the C key.