When the scale shows <Add> or <Sub>:
• If the sample weight is less than 90 times the readability (90 x d), the
display shows <Add> and the unit weight cannot be updated.
In this case, place as many pieces on the weighing plate until the
message is no longer displayed or select a higher reference piece
count (see chapter 10.2).
• If the additional reference piece count is more than twice the initial
reference piece count, the display will show <Sub> and the piece
weight cannot be updated. In this case, reduce the number of pieces
of the additionally placed sample.

16.11 Multi-range mode

The models FEJ 62K-4D and FEJ 62K-4DM are set as multi-range scales 6.2 kg =
0.1 g and 62 kg= 1 g readability as standard. The scales can also be set as single-
range scales with a readability of 1 g.
This function is only available for the weighing systems FES 62K-4D and
FEJ 62K-4DM
61
must not be more than twice as large as the initially
selected reference number of pieces → Example
Selected = 10 pieces, additional reference pieces =
20 pieces or less)
 Press the [F4] key to save the reference weight.
 Use
[F1]
and
<1 APPLICATIONS>.
 Press [F4] key to go to the application settings.
 Navigate to <1D RANGE MODE> with the keys [F1]
to [F4]
 Press [F4] key to make settings
 Select the desired setting with [F1] and [F2].
SGL
Single-range mode
DBL
Multi-range mode
 Press [F4] key to confirm selection.
[F2]
keys
to
TFES_TFEJ-BA-e-2210
navigate
to