18.6 Readability
The greater the readability is set, the less the scale is affected by external influences.
In addition, the scale value stabilises more quickly.
•
This function is not available for the FES 62K-4D and FEJ 62K-
4DM weighing systems.
•
For the FEJ 17K-4M and FEJ 33K-4M weighing systems, the
readability is set to <1> (default) and cannot be changed when
using the "carat" weighing unit
18.7 Adjustment with internal weight at power-up
This function is only available for the FEJ weighing system.
18.7.1 Activate internal adjustment at power-up
69
Open <64 SCALE MANAGE> (see chapter 18.1).
Navigate to <64A READABILIT> with the [F1] and
[F2] keys.
Press [F4] key to make settings
Use [F1] and [F2] to select the desired setting:
1
1 (default)
2
2
5
5
10
10
Press [F4] key to confirm selection.
Open <64 SCALE MANAGE> (see chapter 18.1).
Use [F1] and [F2] to navigate to <64B START
CAL>.
Press [F4] key to make settings
Use [F1] and [F2] to select the desired setting:
OFF
Disabled
FORCE
Always adjust the scale at power-up
Ask if adjustment needs to be performed each time the
SELEC
scale is switched on
Press [F4] key to confirm selection.