Percent Weighing; Selecting The Percent Weighing Function - KERN FES 17K-4 Operating Instructions Manual

Precision balance
Table of Contents

Advertisement

11 Percent weighing

The percent weighing application allows you to check the weight of a sample in per
cent in relation to a reference weight.
Sample in per cent, relative to a reference weight.
• If necessary, set to zero (see chapter 9.2) or tare (see chapter 9.3)
before weighing.
• The readability of the balance automatically adjusts to the reference
weight:
Readability in %
The reference weight can be recorded in two ways:
• Actual value setting method: Weighing the reference weight
• Numeric entering of the reference weight

11.1 Selecting the percent weighing function

If data on a previous reference weight is available:
If no data on a previous reference weight are available or value has not been
stored:
37
Weight range of the reference weight
1
10 g <=
0.1
100 g <=
0.01
1000 g <=
 Select weighing mode <PCNT> (for selection see
chapter 9.5)
Query whether the last reference weight is to be used
again:
 Select the desired setting with [F3] and [F4].
NO
Do not import value
YES
Accept value
 Display <SET 100%> appears
 Select the actual value setting method with the [F3]
key or go to the numerical input of the reference
weight with the [F4] key.
onW
Actual value setting method
NUM
Numeric entering of the reference weight
Reference weight
Reference weight
Reference weight
TFES_TFEJ-BA-e-2210
< 100 g
< 1000 g

Advertisement

Table of Contents
loading

Table of Contents