Response Formats; A00/Exx Format - KERN FES 17K-4 Operating Instructions Manual

Precision balance
Table of Contents

Advertisement

• C3 can contain a maximum of 10 numeric characters (including
polarity +/-, comma and dot).
• Be careful not to enter any weighing units (e.g.,).
• Entries are only accepted by the scale if it is in a weighing application
(simple weighing, piece counting, percentage weighing, coefficient
multiplication). If the scale is not in a weighing application, an incorrect
response will be shown.
• If invalid entries are made, the scale responds with an incorrect
answer.
• Observe the readability of the scale when entering tolerance values or
PRE-TARE values, The entry will not be accepted if the value entered
exceeds the readability range.
Set tolerance values:
Code
C1
C2
(C1)
L
A
4CH
L
B
4CH
L
C
4CH
PRE-TARE value:
Code
C1
C2
(C1)
P
T
50H
• For
<321 PRESET 1> and executed by the scale.
• If a PRE-TARE value of 0 is entered, PRE-TARE is aborted
Time setting of the interval output:
Code
C1
C2
(C1)
I
A
49H

21.6 Response formats

21.6.1 A00/Exx Format

Consisting of 5 characters, including the end characters (CR= 0DH, LF= 0AH)*.
1
2
3
4
A1
A2
A3
CR
* End characters: CR = paragraph, LF = line
TFES_TFEJ-BA-e-2210
Code
(C2)
Set lower tolerance
41H
limit
Set upper tolerance
42H
limit
43H
Set reference value
Code
(C2)
54H
Set PRE-TARE value
normal
responses,
Code
(C2)
Set time of interval output
41H
(hh,mm,ss).
5
LF
Significance
Significance
the
PRE-TARE
Significance
C3
Numeric value
Numeric value
Numeric value
C3
Numeric value
value
is
stored
C3
Numeric value
in
90