Data Description - KERN FES 17K-4 Operating Instructions Manual

Precision balance
Table of Contents

Advertisement

21.4.2 Data description

Status of data:
S1 = 1 character
S1
Code
Sp
20H
*
2AH
*Sp = space
Result of the tolerance range weighing:
C1 = 1 character
C1
Code
Sp
20H
H
48H
L
4CH
*Sp = space
Data type:
T1 – T6 = 1 - 6 characters
T1
T2
T3
T4
Sp
Sp
Sp
Sp
N
Sp
Sp
Sp
P
T
Sp
Sp
T
Sp
Sp
Sp
T
O
T
A
G
Sp
Sp
Sp
U
N
I
T
*Sp = space
Numeric data:
D1 – D12: 1 – 12 characters
D1 – D12 Code
+
2BH
-
2DH
30H – 39H
0 - 9
.
2EH
TFES_TFEJ-BA-e-2210
Significance
Data stable
Data not stable
Significance
Within tolerance range (OK)
Upper tolerance limit exceeded (HIGH)
Below lower tolerance limit (LOW)
T5
T6
T1
Sp
Sp
20H
Sp
Sp
4EH
Sp
Sp
50H
Sp
Sp
54H
L
Sp
54H
Sp
Sp
47H
Sp
Sp
55H
Significance
Zero or positive values
Negative values
Numbers 0 to 9
0 also used for zero padding
Decimal point
Code
T2
T3
T4
T5
20H
20H
20H
20H
20H
20H
20H
20H
54H
20H
20H
20H
20H
20H
20H
20H
4FH
54H
41H
4CH
20H
20H
20H
20H
4EH
49H
54H
20H
Significance
T6
Net weight <41A
20H
STATUS>: <OFF>
Net weight <41A
20H
STATUS>: <ON>
20H
PRE-TARE value
20H
Tare weight value
20H
Total
20H
Gross weight
20H
Piece weight
86