19.2.1 Performing the internal adjustment
...
19.2.2 Adjustment test with internal weight
19.3 Setting the adjustment reminder
The adjustment reminder triggers an alarm when the scale needs a new adjustment.
This is the case if the scale has not been switched on for a long time, if the
temperature or atmospheric pressure is different, or if the last adjustment was made
a long time ago.
This function is only available for the FES weighing system.
77
Open <63 MAINTENANCE> (see chapter 19).
Use [F1] and [F2] keys to navigate to <633 INT
CAL>.
Press [F4] key to start internal adjustment.
Various messages are shown on the display one
after another: <buSY> → <Ch 0> → <Ch FS>
When <633 INT CAL> is displayed again, the
internal adjustment is complete.
Open <63 MAINTENANCE> (see chapter 19).
Use [F1] and [F2] keys to navigate to <634 INT
SPAN TEST>.
Press the [F4] key to start the internal adjustment
test.
Various messages are subsequently shown on the
display: <buSY> → <Ch 0> → <Ch FS> →
<INSTRUMENTAL ER>
Press the [F4] key when <INSTRUMENTAL ER> is
displayed.
When <634 INT SPAN TEST> is displayed again,
the internal adjustment test is complete.
TFES_TFEJ-BA-e-2210