23.1 Error messages

Error message
OVER ERROR
UNDER ERROR
DISPLAY ERROR
LOWER ERROR
ERR001 to
ERR099
ERR703
ERR705
ERR706
ERR709
ERR710
ERR711
TFES_TFEJ-BA-e-2210
Description
Maximum weighing range exceeded
The total exceeds the maximum
number of characters on the display
Negative load is below the minimum
weighing range
The total exceeds the maximum
number of characters on the display
The piece/reference weight in
counting/per cent mode is below the
minimum weighing range
System error
A key was pressed during power-up
If no key was pressed, there may be a
hardware fault.
Initial zeroing error during power-up
due to unstable load
Load is outside the initial zeroing range
Unstable weight value during zeroing
or taring
Timeout during adjustment
Possible causes / repair
Split sample and weigh
individually
Use lighter tare weight
Delete the result of the
calculation and perform the
calculation again (pay
attention to the display)
Weighing plate or weighing
plate carrier incorrectly
adjusted
Check whether the scale is
touching other objects
Only use the intended
weighing plate and
weighing plate carrier
Delete the result of the
calculation and perform the
calculation again (pay
attention to the display)
Select samples whose
piece/reference weight is
above the minimum
weighing range
Note error code and notify
dealer
No key pressed while the
scale is performing the
power-up procedure
Weighing plate or weighing
plate carrier incorrectly
adjusted
Check whether the scale is
touching other objects
Check for vibrations or
draughts
Do not load the scale
during power up
Weighing plate or weighing
plate carrier incorrectly
adjusted
Check whether the scale is
98