2 Technical data 2.1 KERN 572 KERN 572-30 572-31 572-32 572-33 572-35 Readability (d) 0.001 g 0.001 g 0.001 g 0.01 g 0.01 g Weighing range (max) 161 g 241 g 301 g 1610 g 2410 g Taring range (subtractive)
2.2 KERN 573 KERN 573-46 Accuracy class Readability (d) 0.1 g Verification value (e) Weighing range (max) 6500 g Minimum load (Min) Taring range (subtractive) 6500 g Reproducibility 0.1 g Linearity ±0.2 g Smallest piece weight 0.3 g Recommended adjusting weight F2 (not supplied)
2.3 KERN KB KERN KB 120-3 KB 200-3 KB 1200-2 KB 2000-2 KB 10000-1 Readability (d) 0.001 g 0.001 g 0.01 g 0.01 g 0.1 g Weighing range (max) 121 g 201 g 1210 g 2010 g 10100 g Taring range (subtractive)
2.5 KERN DS KERN DS 5K0.05S DS 8K0.05 DS 10K0.1S DS 16K0.1 DS 20K0.1 Readability (d) 0.05 g 0.05 g 0.1 g 0.1 g 0.2 g Weighing range (max) 5100 g 8100 g 10100 g 16100 g 20100 g Taring range (subtrac-...
2.6 KERN PCB KERN PCB 200-3B PCB 240-3B PCB 2000-2B PCB 2400-2B PCB 10K0.05B Readability (d) 0.001 g 0.001 g 0.01 g 0.01 g 0.05 g Weighing range (max) 201 g 241 g 2010 g 2410 g 10.100 g Taring range (subtrac-...
2.7 KERN FCB KERN FCB 12K0.1B FCB 24K0.2B Readability (d) 0.1 g 0.2 g Weighing range (max) 12100 g 24100 g Taring range (subtrac- 12100 g 24100 g tive) Reproducibility 0.1 g 0.2 g Linearity ± 0.3 g 0.6 g Smallest piece weight 0.1 g...
For this purpose, the answerable user must define a suit- able interval as well as the nature and scope of this check. Information is available on KERN’s home page (www.kern-sohn.com) with regard to the monitoring of balance test substances and the test weights required for this.
Observe the information in the operating instructions Please read the operating instructions carefully before erecting and commissioning, even if you already have experience with KERN balances. Staff training The device may only be operated and looked after by trained members of staff.
Electric power supply is by means of the external mains supply circuit. The printed voltage level must comply with the local voltage. Only use original KERN mains supply circuits. The use of other makes is subject to approval by Kern. Connecting peripheral equipment The balance must be disconnected from the mains before connecting or disconnecting additional equipment (printer, PC) to or from the data interface.
6.7.1 Adjusting for verification (KERN 573) General Information: Prior to the procedure of verification the balance has to be adjusted. Remark: The adjusting is only possible when not being blocked by the adjusting switch. The switch for this step is located at the bottom of the housing between the two turning feet.
7 Operation 7.1 Display control panel Keyboard ON / OFF Print weighing result in MODE: No / descending in % and counting mode: Form reference in MODE: Yes / ascending MODE configuration (see mode structure diagram page 2) Formula weighing in MODE: to left Change g-pieces Unit changeover...
7.2.6 Adjust 7.2.7 Preselect Tare 7.2.8 Auto Tare 7.2.9 Speed 7.2.10 Auto Off 7.2.11 Variable Factor 7.2.12 Preselect Unit 572/573/KB/DS/FKB/FCB/PCB-BA-e-0853...
Baud rate adjustable to 2400, 4800 and 9600 baud (default), 19200 baud Sup-D- 9 pol. is necessary . When working with an interface correct operation is secured only if the corresponding KERN-interface- cable (max. 2m) is used. Description of the jack Sup-D- 9 pol.
7.4 RS 232 C Data output via interface RS 232 C RS 232 C Data output via interface RS 232 C General information As a condition for the data transfer between the balance and a peripheral device (for instance printer, PC ...) both devise have to be set on the same interface parameter (for instance baud rate, parity ...).
7.5 Printer With the serial data output RS 232 a printer can be connected. The printout shows the weight in grams. When the counting mode is selected the number of pieces or the weight is printed. When the percent mode is selected, the percentage or the weight will be printed. Press The PRINT-key to print weighing results.
Loose sample residue/powder can be removed carefully using a brush or hand vacuum cleaner. Remove any spilt material to be weighed immediately. 8.2 Maintenance, upkeep The device may only be opened by trained service engineers authorised by KERN. Disconnect from the mains supply before opening. 8.3 Disposal The operating company shall dispose of the packaging and the device in compliance with the valid national or regional law of the operating location.