Functional Tests

Frequency Response (Flatness)

Frequency Response (Flatness)
Test Limits
20 Hz to 3.0 GHz
3.0 GHz to 7.5 GHz
The frequency response test measures the signal analyzer's amplitude error as a
function of the tuned frequency. Measurements are made ranging from 50 MHz to
the maximum frequency range of your analyzer. The signal source amplitude is
measured with a power meter to eliminate error due to source flatness. The
measured value is normalized to 50 MHz.
Item
Adapter, Type N(m) to 3.5
mm(m)
Adapter,
3.5 mm(f) to 3.5 mm(f)
Cable, 3.5 mm, 1 meter
Cables, BNC 120 cm
Power Meter
Power Sensor
Power Splitter
Synthesized Sweeper
304
Frequency Range
Critical Specifications
(for this test)
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.4:1
Frequency: 10 MHz
Compatible with power sensor
Frequency Range: 50 MHz to 26.5 GHz
Frequency Range: 50 MHz to 26.5 GHz
Tracking between ports: < 0.25 dB
Frequency Range: 50 MHz to 26 GHz
Limit Relative to 50 MHz
±1.5 dB
±2.5 dB
Recommended
Agilent Model
1250-1743
83059B
11500E
10503A
E4418B
8485A
11667B
PSG