Scan Probe Environmental Requirements

Operation:10° to 40° C
Storage: 10° to 70° C
NOTE:
Temperature in degrees C. Conversion to Degrees F = (Degrees C * (9/5) + 32).
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF
NOTICE
-20 TO + 50 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
2-2-5

Time and Manpower Requirements

Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six
weeks before delivery, for enough time to make necessary changes.
Have two people available to deliver and unpack the Vivid P3.
CAUTION
Attempts to move the unit considerable distances or on an incline by one person could result
in injury or damage or both.