Page 2
The information in this manual applies only to MAC™2000. Due to continuing product innovation, specifications in this manual are subject to change without notice. MAC is a trademark owned by GE Medical Systems Information Technologies, Inc., a General Electric Company going to market as GE Healthcare. All other trademarks contained herein are the property of their respective owners.
GE Healthcare. Use this manual as your guide for the GE Healthcare supplies and accessories that have proven compatibility with your device. From electrodes, cables, and leadwires...
To ensure patient safety, use only supplies and accessories manufactured or validated by GE Healthcare and recommended for use with this device. Parts, supplies, and accessories must meet the requirements of the applicable safety standards.
Page 12
Supplies and Accessories Leadwires (cont'd.) Description Part Number 38401817 ECG Leadwire Set, Banana, 10/set, AHA, mixed lengths 420101-002 CAM 14 ECG Leadwire Set, Universal, mixed length, 10/set Leadwire Accessories Part Number Description 401089-003 Leadwire Organizers, 4 3-L Separators, 2 4-L Separators 414763-002 Multi-Link Leadwire Separator, 3-L, 3/pack Cables...
Page 13
Supplies and Accessories ECG Adapters and Connectors (cont'd.) Description Part Number 900178-102 CAM 14 ECG Connector Set, Mactrode, 10/set, IEC 900178-103 CAM 14 ECG Connector Set, Grabber, 10/set, IEC KISS Accessories Description Part Number 2024037-001 KISS ES600 Electrodes & Leadwires, Standard, 10/set, mixed lengths 2024039-001 KISS Upgrade Electrodes &...
Page 14
Supplies and Accessories KISS Accessories (cont'd.) Description Part Number 30344439 Leadwire, KISS, AHA, 127 cm (50 in), V6 30344440 Leadwire, KISS, AHA, 147 cm (57 in), LL 30344441 Leadwire, KISS, AHA, 147 cm (57 in), LA 30344442 Leadwire, KISS, AHA, 147 cm (57 in), RA 30344443 Leadwire, KISS, AHA, 147 cm (57 in), RL 38401584...
Page 15
Supplies and Accessories Barcode Accessories (cont'd.) Description Part Number 2035979-018 Barcode Scanner — Portuguese 2035979-020 Barcode Scanner — Finnish Storage Accessories Description Part Number 2073012-001 CARD SECURE DIGITAL - 4 GB Power Cords Description Part Number 401855-001 EURO - PWR SPLY CRD RA CONT EURO 10A 250V 2.5M 401855-002 British - PWR SPLY CRD RA BRITSH 10A 250V 2.5M FSD 401855-003...
Page 16
Supplies and Accessories 2053535-004D MAC™2000 ECG Analysis System...
Maintenance This chapter provides information about the maintenance of the accessories used with your system. Cleaning, Disinfecting, and Storing ECG Cables and Leadwires In addition to keeping the system clean and in good repair, it is important to keep the cables and leadwires clean and disinfected.
Maintenance • Ketone • Betadine • Alcohol-based cleaning agents • Sodium salts • Never autoclave or steam clean the device, cables, or leadwires. • Do not use any of the accessories until thoroughly dry. • Do NOT immerse either end of a cable or leadwire connector in any solution. Immersing or soaking the connector ends may corrode metal contact ends and affect signal quality.
Maintenance • Clorox® Wipes (they do not contain bleach) • Over-the-counter detergents (such as Fantastic®, Tilex®, and so on) Wipe the cables and leadwires with a dry, clean cloth or paper towel and let them air dry. Disinfecting the Cables and Leadwires Use the following procedure to disinfect the cables and leadwires: Clean and dry the cables and leadwires before disinfecting them.
Maintenance Cleaning and Disinfecting Reusable Electrodes Use the following procedure to clean and disinfect the reusable electrodes: Use warm water and a small brush to remove any cream or gel from the electrode. Do not use pointed or sharp objects for cleaning. Disinfect the electrodes with alcohol-free disinfectant.
Maintenance Paper Maintenance For the proper handling of the device's thermal writer, you need to know how to do the following: • Replace paper • Store thermal paper Replacing Paper Use the following procedure to add or replace paper: Press the push button on the top of the device (1) to open the printer door (2). Extend the top sheet of the pack of paper and insert the pack into the paper compartment (3).
Storing Thermal Paper When imaged and stored properly, ECG tracings resist fading for several years. If your retention requirements exceed five years, consider using GE Healthcare Archivist paper. To ensure the tracing is imaged properly, the device must be maintained in accordance with its service manuals and technical memoranda.
Page 23
Index accessories 6 GE Healthcare Common Documentation Library (CDL) 2 manuals 2 barcode accessories 14 gels 9 guidelines when cleaning and disinfecting 17 cables 12 cleaning and disinfecting 18 storing 19 cleaning KISS guidelines 17 accessories 13 solutions not to use 17...
Page 24
paper replacing 21 storing 22 Paper 9 part number document 2 patient vicinity definition 7 power cords 15 regulatory information 6 replacing leadwire adapters 20 revision history 2 safety information 6 solutions not to use when cleaning or disinfecting 17 sprays 9 sterilizing electrodes 20...
Page 26
+49 761 45 43 -233 Shanghai, People's Republic of China 201203 Fax: +1 414 355 3790 Tel: +86 21 3877 7888 Fax: +86 21 3877 7451 GE Medical Systems Information Technologies, Inc., a General Electric Company, going to market as GE Healthcare. www.gehealthcare.com 0459...