Communication with the DMP
RS‐232 interface
16
Serial interface characteristics (default settings)
Word length
Stop bit
Parity
Baud rate
1)
Factory setting
2)
Can only be set with command BDR
Baud rate, parity and stop bits can be set in the settings menu or by
command.
3.3.1
Activation of the RS‐232 interface
The HBM interpreter is activated by the control character:
S CTRL B (STX) - Computer operation
S CTRL R (DC2) - Computer operation
Entering one of these control characters puts the device into the remote
control state "Computer Control" so that, apart from the display functions of
the display, it can still be operated but no longer set.
With the RS-232-C interface, all information generated is immediately output
once fully present in the output buffer.
You can deactivate the remote control state with the following commands:
DCL or
CTRL A (SOH) or
with the RES and RST commands
Important
An RS232 interface can only be operated with the DMP41 via an adapter
(USB -> RS232).
A USB-RS232 adapter is included in the scope of delivery of the DMP41
(1-KAB297).
A03474_04_E00_00 HBM: public
8 bit
1
Even
300, 600, 1200, 2400 2) , 4800 2) , 9600 1) ,
19 200, 38400, 576000, 115200