HBM DMP41 Interface Description page 69

Digital precision measuring device
Hide thumbs Also See for DMP41:
Table of Contents

Advertisement

IAD?
DMP41
Acknowledgement
0
?
Example:
IAD2,10000,3,4(x)
0(y)
Set display adaptation for measuring range 2
Display full scale 10.000 with step 0.010
Important
Only the range momentarily selected with ASA can be input in measuring
range 1; the decimal places can vary between 3 and 6.
Permissible, e.g.:
Indication Adaptation Query
Output display full scale, decimal point, step
Syntax:
IAD? p1(x)
Parameter:
p1
1
from measuring range 1
2
from measuring range 2
Effect:
Output of display full scale, decimal point, step.
Response:
q1,q2,q3,q4(y)
Parameters see IAD command
Example:
IAD?2(x)
2,10000,3,1(y)
Display full scale in measuring range 2 is 10.000 with step
0.001.
A03474_04_E00_00 HBM: public
Command set of the HBM interpreter
Significance
Command executed
Error
IAD1,250000,6,2
IAD1,50000,4,1
Status
Amplifier functions
69