Command set of the HBM interpreter
Amplifier functions
LTB
LTB?
70
Linearization Table
Linearization of transducer characteristic curve
Syntax:
LTB n,x
Parameters:
n=2 ...11
x=Numbers in mV/V (Measuring range 1)
y=Numbers with actual unit from measuring range 2
Effect:
This command can be used to correct known linearity devia
tions of the transducer by specifying up to 11 points of the
characteristic curve.
Important
Value pairs are sorted according to X and checked that they increase or
decrease monotonically.
"0" = OK
"?" = NOK
Linearization Table Query
Output actual linearization curve
Syntax:
LTB?(x)
Parameter:
none
Effect:
The actual characteristic curve points are output.
Important
After the characteristic curve points are entered, the display adaptation
(command "IAD") of the measuring range 2 is automatically adapted. As the
linearization table is input in mV/V, a correction after a measuring range
change is not necessary.
Values that lie outside the momentarily selected measuring range can also
be input.
A03474_04_E00_00 HBM: public
,y
x
,y
(x)
1
1 ....
n
n
DMP41