HBM DMP41 Interface Description page 70

Digital precision measuring device
Hide thumbs Also See for DMP41:
Table of Contents

Advertisement

Command set of the HBM interpreter
Amplifier functions
LTB
LTB?
70
Linearization Table
Linearization of transducer characteristic curve
Syntax:
LTB n,x
Parameters:
n=2 ...11
x=Numbers in mV/V (Measuring range 1)
y=Numbers with actual unit from measuring range 2
Effect:
This command can be used to correct known linearity devia­
tions of the transducer by specifying up to 11 points of the
characteristic curve.
Important
Value pairs are sorted according to X and checked that they increase or
decrease monotonically.
"0" = OK
"?" = NOK
Linearization Table Query
Output actual linearization curve
Syntax:
LTB?(x)
Parameter:
none
Effect:
The actual characteristic curve points are output.
Important
After the characteristic curve points are entered, the display adaptation
(command "IAD") of the measuring range 2 is automatically adapted. As the
linearization table is input in mV/V, a correction after a measuring range
change is not necessary.
Values that lie outside the momentarily selected measuring range can also
be input.
A03474_04_E00_00 HBM: public
,y
x
,y
(x)
1
1 ....
n
n
DMP41