Command set of the HBM interpreter

Amplifier functions

COF
56
5.4
Amplifier functions
5.4.1
Output format, measured value output
Change Output Format
Change measured value output format
Syntax:
COF p1(x)
Parameter:
p1
0
Measured value, channel, status (ASCII format)
1
2
Binary measured value output 4 byte (MSB XX XX LSB)
3
Binary measured value output 4 byte (LSB XX XX MSB)
4
Binary measured value output 2 byte (MSB LSB)
5
Binary measured value output 2 byte (LSB MSB)
4 byte output:
MSB
LSB
1
2
3
4
3 byte measured value
7680000 = Measuring range full scale (unit)
The output value multiplied by 1/2560 is equivalent to the measured value in
degrees for thermocouples and the units 5C,5F, K.
Sta­
Significanc
tus
e
0
OK
Measureme
nt without
warnings or
error
1
Warn
Filter setting
ing
2
Overflow
A03474_04_E00_00 HBM: public
Measured value output format
Measured value (ASCII format)
1 byte status
Bit
Bit
Bit
Bit
Bit
7
6
5
4
3
0
0
0
0
LV
4
0
x
x
1
LV
4
0
x
1
x
LV
4
Bit
Bit
Bit
GUI
2
1
0
message
LV
LV
LV
3
2
1
LV
LV
LV
Change color
3
2
1
of filter text
LV
LV
LV
Display
3
2
1
warning
symbol
DMP41

Advertisement

Table of Contents
loading

Table of Contents