Error Control, Status Register - HBM DMP41 Interface Description

Digital precision measuring device
Hide thumbs Also See for DMP41:
Table of Contents

Advertisement

XST?
DMP41
5.2.3

Error control, status register

Extended Status Query
Extended status query
Syntax:
XST?(x)
Parameter:
none
Effect:
The DMP41 responds with a bit‐coded decimal number.
Bit
0
1
2
3
4
5
6
7
8
9
10
Bit0:
Transducer error
Bit1:
Calibration error
Bit2:
No calibration carried out
Bit3:
Timeout output stage
Bit4:
Amplifier overloaded
Bit5:
Sense lead (2-2') interrupted
Bit6:
Sense lead (3-3') interrupted
Bit7:
Calibration running
Bit8:
Background calibration running
Bit9:
Filter settling time
Bit10:
Background calibration error
Bit11:
No 0‐wire detected
Bit12:
No 1‐wire detected
Bit13:
Converters (internal) not synchronous
Bit14:
Initialization (new transducer)
Bit15:
Modulator overloaded
XST? outputs the sum of all relevant bits.
Response 258 means: 256 (calibration) and 2 (calibration error). This is a
typical response after a channel change and ongoing initial calibration.
A03474_04_E00_00 HBM: public
Command set of the HBM interpreter
Hex
0001
0002
0004
0008
0010
0020
0040
0080
0100
0200
0400
Communication
Decimal
1
2
4
8
16
32
64
128
256
512
1024