Command set of the HBM interpreter
Amplifier settings
CMR?
48
p1
Measuring range code number
1
Measuring range (Range) 1
2
Measuring range (Range) 2
Effect:
The required measuring range is switched on.
Response:
Acknowledge-
ment
0
Command executed
?
Example:
CMR2(x)
0(y)
Measuring range 2 set.
Change Measuring Range Query
Output of measuring range
Syntax:
CMR?(x)
Parameter:
none
Effect:
The selected measuring range is output.
Response:
q1(y)
q1
Measuring range code number
1
Measuring range (Range) 1
2
Measuring range (Range) 2
Example:
CMR?(x)
2(y)
Measuring range 2 set.
A03474_04_E00_00 HBM: public