Command set of the HBM interpreter
Amplifier settings
ASA?
40
Example:
ASA3,1(x)
0(y)
The amplifier is set to 10 V bridge excitation voltage,
2.5 mV/V sensitivity and Shunt 'OFF'.
Important
This command triggers a calibration procedure in the measuring amplifier.
Amplifier Sensor Adaptation Query
Output bridge excitation voltage and transducer type
Syntax:
ASA? p1(x)
Parameter:
p1
0
Output bridge excitation voltage and input sensitivity settings
1
Output table of possible bridge excitation voltage and input
Effect:
The amplifier outputs the set bridge excitation voltage and
input sensitivity (p1=0) or their possible values (p1=1).
Response:
ASA?0(x)
q1,q2(y)
q1
Bridge excitation voltage
q2
Example:
ASA?1(x)
"
02.505.010.0",
Response:
ASA?1(x)
q1,q2(y)
Table of possible settings
q1
Possible bridge excitation voltage
q2
Possible sensitivities
Bridge excitation voltages
A03474_04_E00_00 HBM: public
sensitivity settings
Input sensitivity
"
"
123
DMP41