Command set of the HBM interpreter
Amplifier settings
AFS
AFS?
42
Parameter:
none
Effect:
Type of amplifier input signal is output.
Response:
q1(y)
q1
Input signal source of the amplifier
0
Amplifier input is switched to ZERO.
1
Amplifier input is switched to CAL.
2
Amplifier input is switched to MEAS.
Example:
ASS?(x)
2(y)
Amplifier input is switched to MEAS (measurement signal).
5.3.2

Filter setting

Amplifier Filtering Select
Filter switchover (fc 1/2)
Syntax:
AFS p1(x)
Parameter:
p1
Filter code number
1
2
Effect:
Switchover between two filters (see also ASF).
Response:
Acknowledge-
ment
0
Command executed
?
Example:
AFS 2 (x)
0(y)
Filter fc2 is switched on.
Amplifier Filtering Select Query
Output filter settings
Syntax:
AFS?(x)
Parameter:
none
A03474_04_E00_00 HBM: public
fc1
fc2
Significance