HBM DMP41 Interface Description page 68

Digital precision measuring device
Hide thumbs Also See for DMP41:
Table of Contents

Advertisement

Command set of the HBM interpreter
Amplifier functions
IAD
68
"
V_ _ _G_ _ _KG_ _T_ _ _KT_ _TONSLBS_N_ _ _KN_
_BAR_mBARPA_
PAS_HPASKPASPSI_UM_ _MM_ _CM_ _M_ _ _INCHNM_ _FTL­
BINLB
UM/MM/S_M/SSp/o_p/
Output table of possible units.
This string contains the table of possible units for this ampli­
fier. Each element in the table is 4 characters long.
Indication Adaptation
Input, display full scale, decimal point, step
Syntax:
IAD p1,p2,p3,p4(x)
Parameter:
p1
p2
Display full scale without decimal point
p3
Decimal point (number of decimal places)
p4
Step (= minimum digit jump) (digit)
The step p4 can have the following values:
p4
Step
1
1
2
2
3
5
4
10
5
20
6
50
7
100
8
200
9
500
10
1000
Effect:
This command is used to input display adaptation values.
Response:
A03474_04_E00_00 HBM: public
"
PPM_
(y)
00
Range 1 or 2