HBM DMP41 Interface Description page 64

Digital precision measuring device
Hide thumbs Also See for DMP41:
Table of Contents

Advertisement

Command set of the HBM interpreter
Amplifier functions
STP
TEX
64
Example:
Measurement of transducer temperature with 1-wire tem­
perature sensor
MEV?1(x)
2650.0
o
26,50
C, Status OK
MEV?15 (15 = Channel mask)
2650,0,2850,0,2670,0,0,5
Channel 1: 26.5
Channel 2: 28.5
Channel 3: 26.7
Channel 4: no temperature sensor
Delivers temperature and temperature status of all 4 tem­
perature sensors (4 temperature sensors, also for
DMP41-T2).
Status table:
Status
0
1
2
3
4
5
6
7
8
Stop
Stop measured value output
Syntax:
STP(x)
Parameter:
none
Effect:
The measured value output started with MSV? is stopped.
Response:
none
Example:
STP(x)
Define Terminator
Define measured value separator
Syntax:
TEX p1,p2(x)
A03474_04_E00_00 HBM: public
o
C, Status OK
o
C, Status OK
o
C, Status OK
Meaning
TEMP_NOERROR
TEMP_NEVER_MEASURED
TEMP_SHORT_CIRCUIT
TEMP_NO_PRESENCE_PULSE
TEMP_MANY_DEVICES
TEMP_NO_TEMP_SENSOR
TEMP_SEARCH_ROM_FAILED
TEMP_CONVERT_FAILED
TEMP_READING_FAILED
DMP41

Advertisement

Table of Contents
loading

Table of Contents