HBM DMP41 Interface Description page 47

Digital precision measuring device
Hide thumbs Also See for DMP41:
Table of Contents

Advertisement

CDW?
CMR
DMP41
CDW?: If correspondingly programmed, the output value is inverted
See also SGN
A zero value that is more "10.1 mV/V will not be accepted.
Calibration Dead Weight Query
Output of zero value
Syntax:
CDW? p1(x)
Parameter:
p1
0
1
Actual absolute value in ADU
10
11
12
Zero value in scaled units
Effect:
This command outputs the zero value from the memory or
the actual absolute
Response:
q1(y)
q1
Zero value in (equal to p1) required unit
7680000 ADU units correspond to the currently set measuring range full
scale (Range 1) 2.5 mV/V, 5 mV/V or 10 mV/V.
Example 1:
CDW?1(x)
10000(y)
The current ABS measured value is output.
This signal is now set to zero with CDW 10 000(x).
CDW?11(x)
2.5(y)
Zero value output in mV/V
CDW?0(x)
7680000(y)
Zero value output in ADU
Change Measuring Range
Measuring range switchover (Range 1/2)
Syntax:
CMR p1(x)
Parameter:
A03474_04_E00_00 HBM: public
Command set of the HBM interpreter
Return value
Zero value in ADU
Zero value in ADU
Zero value in mV/V
value
.
Amplifier settings
47