Command set of the HBM interpreter
Amplifier settings
46
Effect:
The input value is saved in the amplifier zero memory.
Response:
Acknowledge-
ment
0
Command executed
?
Example 1:
Start zero setting
CDW(x)
0(y)
Example 2:
CDW10000(x)
0(y)
Information
If the value read out with CDW?1 (absolute value) is sent for p1, the
momentarily present gross measurement signal is set to zero.
CDW7680000(x)
0(y)
Zero value set to 7680000 ADU
CDW2.5,11(x)
0(y)
Zero value set to 2.5 mV/V
CDW780.75,12(x)
0(y)
Zero value set to 780.75 in scaled unit
Important
It is still possible to set a basic misalignment to zero by calculating the zero
value to be input using the following calculation:
Zero value (ADU units) =
CDW: If correspondingly programmed, the input value is inverted
A03474_04_E00_00 HBM: public
Significance
Error
7680000 x basic misalignment (mV/V)
Measuring range full scale (mV/V)