HBM DMP41 Interface Description page 66

Digital precision measuring device
Hide thumbs Also See for DMP41:
Table of Contents

Advertisement

Command set of the HBM interpreter
Amplifier functions
ENU
ENU?
66
5.4.2
Display functions
Important
The commands in this "Display functions" chapter have an effect on the
measured values shown in the display.
Engineering Unit
Input of unit
Syntax:
ENU p1,p2(x)
Parameter:
p1
Measuring range to which the unit applies
1 = Measuring range 1; 2 = Measuring range 2
p2
The required unit as a string
4 characters
Effect:
Unit is set with p2.
Response:
Acknowledge-
ment
0
Command executed
?
Example:
ENU 2,"KG_ _"(x)
0(y)
Engineering Unit Query
Output of unit
Syntax:
ENU? p1(x)
Parameter:
p1
0
Unit of the momentarily set measuring range
1
2
3
Effect:
The selected unit is output.
A03474_04_E00_00 HBM: public
Significance
Error
Measuring range 1 unit
Measuring range 2 unit
Table of possible units
DMP41