HBM DMP41 Interface Description page 79

Digital precision measuring device
Hide thumbs Also See for DMP41:
Table of Contents

Advertisement

DRS
RS2?
EST?
DMP41
Device Reset
Reset ranges to factory settings
Syntax:
DRSp1(x)
Parameter:
p1
p1
Significance
1
Reset channel settings to factory
settings
2
Reset hardware settings to factory
settings
3
Reset hardware settings to factory
settings
RS232 adapter detected
RS232 adapter detected
Syntax:
RS2?(x)
Effect:
RS232 adapter detection
Response:
q1(y)
q1
1
USB−RS232 adapter was detected
0
No adapter connected or adapter not supported
Extended Status
Syntax:
EST?(x)
Effect:
Returns the extended status once and then sets it to 0.
The extended status provides information about the last
occurred error
(command acknowledged with "?") information
Response:
q1(y)
q1
10003
10004
Too many or too few parameters included (e.g. CHS,1,2,3)
10005
Value of a parameter outside valid limits
10008
Command cannot be executed (e.g. a channel could not be
set to zero because no valid measured value was present)
A03474_04_E00_00 HBM: public
Command set of the HBM interpreter
Range, Filter, Scaling, Name,
Resolution, Zero, MinMax
Password, Brightness, DeviceName
Status
Status
Unknown command was sent
Other
Content
Ethernet, RS232
79

Advertisement

Table of Contents
loading