Command set of the HBM interpreter
Communication
BDR?
34
The RS‐232‐C interface was set to 19200 baud, even parity,
1 stop bit.
Example 2:
The DMP41 is operated via the RS‐485 interface:
BDR4800,0,2(x)
0(y)
The RS‐485 interface was set to 4800 baud, no parity, 2 stop bits.
Important
The response after a BDR command is always output with the changed set­
ting.
Baud Rate Query
Output baud rate of serial interfaces
Syntax:
BDR? p1(x)
Parameter:
p1
Interface whose baud rate is being queried
0
The interface from which the DMP41 is operated
1
RS−232−C interface
2
Effect:
The set baud rate, parity bit, number of stop bits and the ID
of the serial interface are output.
Response:
q1,q2,q3,q4(y)
q1
Baud rate
q2
Parity
q3
Stop Bits
q4
Interface identification
Example 1:
The DMP41 is operated via the RS‐232‐C interface:
BDR?(x)
4800,0,2,1(y)
The RS‐232-C interface is set to 4800 baud, no parity, 2
stop bits.
Example 2:
BDR?2(x)
4800,0,2,2(x)
A03474_04_E00_00 HBM: public
RS−485 interface
DMP41