HBM DMP41 Interface Description page 55

Digital precision measuring device
Hide thumbs Also See for DMP41:
Table of Contents

Advertisement

SLN
SLN?
DMP41
Slot Name
Assign slot name
Syntax:
SLNp1,p2(x)
Parameter:
p1
p1
Slot number (1 ... 100)
1
P2
1
Slot name (string)
Effect:
The channel settings are saved with the TDD command.
The command makes it possible to assign a freely definable
name to each of these slots (1...100).
Beispiel:
Das MGC wird von der IEEE‐Schnittstelle gesteuert
Example:
SLN1,"TestSlot"(x)
0(y)
Assigns Slot 1 the name "TestSlot".
Slot Name Query
Query slot name
Syntax:
SLN?p1(x)
Parameter:
q1
1
Returns the name of the slot. SLN?1 ... SLN?100 are possi­
Effect:
ble
Example:
SLN?1(x)
Master channel
Returns the name of Slot1, e.g. "Master channel"
SLN?0
Outputs a list of all names.
"TestSlot,"2","NameSlot3", ... "Slot name for Slot100"
Response:
String (14 characters)(y)
Example:
IDN?(x)
HBM,DMP41, 4D:5B:B9:02:00:00:1.0.3.2(y)
A03474_04_E00_00 HBM: public
Command set of the HBM interpreter
Status