HBM DMP41 Interface Description page 41

Digital precision measuring device
Hide thumbs Also See for DMP41:
Table of Contents

Advertisement

ASS
ASS?
DMP41
p1
Bridge excitation voltage (V)
02.5
05.0
10.0
The index corresponds to the bridge excitation voltage to be set (each ele­
ment in the table is 4 characters long).
Sensitivity
p2
1
Only 2.5 mV/V possible (10 V excitation)
12
2.5 mV/V or 5 mV/V possible (5 V excitation)
123
2.5 mV/V, 5 mV/V, 10 mV/V possible (2.5 V excitation)
Each element corresponds to the transducer type to be set (see assignment
of code number to transducer type. Each element in the table is 1 character
long).
Amplifier Signal Select
Select amplifier input signal
Syntax:
ASS p1(x)
Parameter:
p1
Input source
0
ZERO Internal zero signal
1
CAL Internal calibration signal
2
MEAS Measurement signal
Effect:
Specification of amplifier input signal.
Response:
Acknowledgement
0
?
Example:
ASS0(x)
0(y)
Amplifier input switched to ZERO (internal zero signal).
Amplifier Signal Select Query
Output of input signal type
Syntax:
ASS?(x)
A03474_04_E00_00 HBM: public
Command set of the HBM interpreter
2.5
5.0
10.0
Sensitivity
Significance
Command executed
Error
Amplifier settings