To Remove And Replace The Front Panel Assembly - Agilent Technologies 16901A Service Manual

Logic analysis system
Hide thumbs Also See for 16901A:
Table of Contents

Advertisement

To remove and replace the front panel assembly

16901A Logic Analysis System Service Guide
1
Perform previous procedures:
"To prepare the instrument for disassembly"
page 46.
"To remove and replace the cover"
Disconnect the following cables:
2
USB cables from the motherboard.
Display cable from the display board (use care when
disconnecting to avoid damage).
Touch screen cable from the touch screen controller
board.
Keyboard cable from the keypad board.
Front panel cable from the backlight inverter board.
Using a Torx T10 screwdriver, remove 4 screws that
3
secure the shroud to the chassis.
4
Remove the shroud.
Replacing Assemblies
on
on page 50.
Shroud
T-10
screws (4)
6
57