Fig. 4-8: Socket X4 - Pin Configuration - Emerson Rosemount Analytical X-STREAM X2 Series Instruction Manual

Gas analyzers
Hide thumbs Also See for Rosemount Analytical X-STREAM X2 Series:
Table of Contents

Advertisement

Instruction Manual
HASX2E-IM-HS
10/2012
Digital inputs
Type:
Electrical specification:
Digital outputs
Type:
Electrical specification:
Note!
Consider the installation notes in section 4.5.
Signal
Input 1
Input 2
Input 3
Input 4
GND for all digital inputs
unused
unused
Output 5 NC
Output 5 NO
Output 5 COM
Output 6, NC
Output 6, NO
Output 6, COM
Output 7, NC
Output 7, NO
Output 7, COM
Output 8, NC
Output 8, NO
Output 8, COM
Fig. 4-8:
Socket X4 - Pin configuration
Emerson Process Management GmbH & Co. OHG
4.6.1 Installation - X-STREAM X2GK
Pin
Pin
1
20
2
21
3
22
4
23
5
24
6
25
7
26
8
27
9
28
10
29
11
30
12
31
13
32
14
33
15
34
16
35
17
36
18
37
19
7 inputs
max. 30 V
, internally limited to 2.3 mA
H signal: min. 4 V;
L signal: max. 3 V
common ground (GND), electrically
isolated from chassis earth connection
9 outputs, dry relay change-over ontacts,
can be used as normally open (NO) or
normally closed (NC)
max. 30 VDC, 1 A, 30 W
Signal
Input 5
Input 6
Input 7
Output 9, NC
Output 9, NO
Output 9, COM
Output 10, NC
Output 10, NO
Output 10, COM
Output 11, NC
Output 11, NO
Note!
Output 11, COM
The configuration illustrated
Output 12, NC
here is that of the first socket,
Output 12, NO
labelled X4.1.
Output 12, COM
Inputs 8-14 and outputs 14-
Output 13, NC
Output 13, NO
22, are on the second socket
Output 13, COM
(X4.2), if installed.
X-STREAM X2
4-13