Instruction Manual
HASX2E-IM-HS
10/2012
start
zero all
procedure
select
channel 1
select all
channels with
the same
zero gas valve
assigned
zero all
channels with
the current
zero gas valve
assigned
is current
channel the
last channel?
increase
channel no.
by 1
is this
Yes
channel
already
zeroed?
Fig. 7-6:
Zero Span All Calibration Procedure Flow Diagram
Emerson Process Management GmbH & Co. OHG
7.4.3 Advanced Calibration
Yes
No
No
X-STREAM X2
start
span all
procedure
select
channel 1
select all
channels with
the same
span gas valve
assigned
span all
channels with
the current
span gas valve
assigned
Yes
is current
channel last
channel?
No
increase
channel no.
by 1
Yes
is this
No
channel
already
spanned?
END
7-29