Emerson Rosemount Analytical X-STREAM X2 Series Instruction Manual page 245

Gas analyzers
Hide thumbs Also See for Rosemount Analytical X-STREAM X2 Series:
Table of Contents

Advertisement

Instruction Manual
HASX2E-IM-HS
10/2012
Setup..
Calibration..
Calibration gases..
Calibration gases
ZeroGas:
SpanGas:
Emerson Process Management GmbH & Co. OHG
7.4.1 Preparing Calibrations
0.0 ppm
500.0 ppm
Before starting calibrations it is required to
tell the instrument the calibration gas concen-
trations.
Starting from the measurement screen press
the DOWN key to open the MAIN MENU,
enter the SETUP-CALIBRATION.. menu and
directly enter the CALIBRATION GASES..
menu.
Multi-channel unit:
Select the channel to be calibrated in the
SELECT COMPONENT menu.
Enter the concentration value for the zero gas
to be used during zero calibration.
Enter the concentration value for the span
gas to be used during span calibration.
Note!
The units for the calibration gases are taken
from the related entry in the display setup
menu.
Multi-channel unit:
Press the  key to enter the SELECT COM-
PONENT menu to change the settings for a
different channel.
When done, press the  key to return to the
CALIBRATION menu.
X-STREAM X2
7-7