Resetting A Calibration - Emerson Rosemount Analytical X-STREAM X2 Series Instruction Manual

Gas analyzers
Hide thumbs Also See for Rosemount Analytical X-STREAM X2 Series:
Table of Contents

Advertisement

X-STREAM X2
7.4.6

Resetting a Calibration

CANCEL calibration!
START calibration!
ZeroGas
CO2.1
RESET calibration..
CO2.1
RESET calibration
Are you sure?
No!
Yes!
7.4.7
Verifying a Calibration
Control..
Zero calibration..
Span calibration..
Adv. calibration..
Apply gas..
7-40
7.4.6 Resetting a Calibration
0.000 ppm
0.200 ppm
0.200 ppm
In case a wrong configuration was detected
after calibration was performed (e.g. wrong
gas connected) there is an option to restore
the last user saved calibration data:
Within the screen where to start the calibrati-
on (either for span or zero) open the second
page:
A new screen appears with the option RESET
CALIBRATION.. Pressing the ENTER key in
this line results in a prompt for confirmation.
Choosing YES! replaces the current calibrati-
on data with the last calibration data, restored
from UserData;
7-54.
For instruments without internal and/or
external valves simply apply either span or
zero calibration gas to the sample gas inlet.
If the calibration still is proper, the reading
on the measurement screen should show the
related value.
For instruments with internal and/or external
valves follow the procedure below:
Starting from the measurement screen press
the DOWN key to open the MAIN MENU,
enter the CONTROL.. menu.
Enter the last line (APPLY GAS..)
Multi-channel unit:
Select the component to be verified in the
SELECT COMPONENT menu.
Emerson Process Management GmbH & Co. OHG
Instruction Manual
HASX2E-IM-HS
10/2012
7.6.2 SAVE-LOAD, page