7.2 General Maintenance Information

Intervals given in the following tables are
based on standard operating conditions
(ambient temperatures +10 ... +40 °C /
+50 ... +104 °F; temperature changes < 10 K
/hr).
Try cleaning contaminated components.
Replace components showing corrosion, or
not passing inspections or tests!
Component
Tubing, flexible
Tubing, stainless steel (SS)
Pressure sensor, pressure switch,
Flowmeter
Pump
Valve block
Flame arrestors
Field housings (IP 66 / NEMA 4X)
Field housings stopping plugs
Field housings cable glands
7-2
7.2 General Maintenance Information
Visual Inspections
Leakage, embrittlement, contamination
Corrosion, contamination
Corrosion, leakage
Fixed screws, swing free to move
Corrosion, leakage
Corrosion, damages, firmly seated
Corrosion, damages on enclosure and
gaskets
Firmly seated
Firmly seated
Instruction Manual
Maintenance intervals must be
shortened for differing opera-
ting conditions, and if aggres-
sive gases are supplied.
Take care of special mainte-
nance instructions in separate
manuals for accessories or
safety equipment, e.g. flame ar-
restors, infallible containments,
etc.
If applicable, consider the ma-
nual addendums for instru-
ments for hazardous areas!
Once a year
Emerson Process Management GmbH & Co. OHG
HASX2E-IM-HS
10/2012