Emerson Rosemount Analytical X-STREAM X2 Series Instruction Manual page 309

Gas analyzers
Hide thumbs Also See for Rosemount Analytical X-STREAM X2 Series:
Table of Contents

Advertisement

Instruction Manual
HASX2E-IM-HS
10/2012
Message
Status level
Explanation
Calibration
Function check
Explanation:
Calibration active
SvcPort > Cf
Function check
Explanation:
SVC port in use
Unlinear
Off-spec
Explanation:
This message does not
appear during operation
modes!
Emerson Process Management GmbH & Co. OHG
8.2.1
Analyzer related messages
Description
Currently a calibration proce-
dure is ongoing
Instrument is currently confi-
gured via service interface
SVC port is in pass through
mode
Instrument has been set to
"unlinear mode" by service
personnel (see "simulation")
X-STREAM X2
Actions
Wait for calibration to finish
Cancel calibration
Wait for data transfer to finish
Switch back to normal com-
munication mode
Switch off diagnostic mode
(Modbus command)
8-5