7.4.1 Preparing Calibrations
Internal
Yes
Yes
Yes
External
Yes
Yes
Yes
Int+Ext
Yes
Yes
Yes
Before starting to assign valves to gases and
channels, you need to check what type of
valves is used (external or internal):
Open the SETUP - INSTALLED OPTIONS..
menu and check the first line ("Valves:")
If this menu (as the example to the left) shows
there are internal valves installed, open the
INTSHS menu (
7.3.1.1.1, page 7-10) to
assign valves.
If the menu (as the example to the left) shows
there are external valves installed, open the
DIGITAL OUTPUTS menu (
page 7-12) to assign valves.
If the menu (as the example to the left) shows
there are internal and external valves in-
stalled, valve assignment has to be done
utilizing both, the INTSHS (
page 7-10) and the DIGITAL OUTPUTS menu
(
7.3.1.1.2 page 7-12).
X-STREAM X2
7.3.1.1.2
7.3.1.1.1
7-11