X-STREAM X2
Signal inputs and outputs
The number of signal outputs actually available
varies according to the unit's configuration.
Analog signals
Relay outputs
Analog signals and relay outputs are
on a shared 25-pin submin socket (X1; Fig. 4-13).
Specification of analog signal outputs:
Specification of relay outputs 1-4:
Electrical specification:
Note!
Consider the installation notes in section 4.5.
Signal
Channel 1, (+) 4 (0)–20 mA
Channel 2, (+) 4 (0)–20 mA
Channel 3, (+) 4 (0)–20 mA
Channel 4, (+) 4 (0)–20 mA
unused
unused
unused
Output1 (Failure) NC
Output1 (Failure) NO
Output1 (Failure) COM
Output3 (Off spec) NC
Output3 (Off spec) NO
Output3 (Off spec) COM
Fig. 4-13: Socket X1 - Analog & Digital Outputs 1-4
4-18
4.6.2 Installation - X-STREAM X2GP
Pin
1
2
3
4
5
6
7
8
9
10
11
12
13
4 (0)–20 mA; burden: R
Dry relay change-over contacts
can be used as NO or NC.
max. 30 VDC, 1 A, 30 W
Pin
Signal
14
Channel 1, GND
15
Channel 2, GND
16
Channel 3, GND
17
Channel 4, GND
18
unused
19
unused
20
Output2 (Maintenance request) NC
21
Output2 (Maintenance request) NO
22
Output2 (Maintenance request) COM
23
Output4 (Function check) NC
24
Output4 (Function check) NO
25
Output4 (Function check) COM
Note!
Configuration of relay contacts
as per standard factory setting (NAMUR
status signals)
Emerson Process Management GmbH & Co. OHG
Instruction Manual
HASX2E-IM-HS
10/2012
≤ 500 Ω