Emerson Rosemount Analytical X-STREAM X2 Series Instruction Manual page 215

Gas analyzers
Hide thumbs Also See for Rosemount Analytical X-STREAM X2 Series:
Table of Contents

Advertisement

Instruction Manual
HASX2E-IM-HS
10/2012
6.2.3.4.4
IntSHS Menu
Setup..
In/Outputs..
IntSHS..
Gas1:
Gas2:
Gas3:
Gas4:
Page 1
Gas5:
Gas6:
Gas7:
Gas8:
Page 2
Pump1:
Pump2:
Page 3
Emerson Process Management GmbH & Co. OHG
6.2 Menu System - IntSHS Menu
V1
V2
V3
V4
Sample
V5
V6
V7
Pump
Off
Note!
This menu appears only when the parameter
in the "Installed Options - Valves" is set to
Internal or IntExt (
This menu configures the optional internal
components for routing gas (valves and
pumps) to be used in autocalibration proce-
dures.
Each available gas inlet ("Gas 1 ... Gas 8")
with a valve connected is assigned a virtual
valve label (V1...V8, Sample). (If the compo-
ents were installed in the factory, the basic
settings will already have been set).
These valves are then assigned a channel
and function (zero or span gas) in the VALVE
ASSIGNMENT MENU (
6-26).
Note!
Depending on the model, 1 or 2 valve blocks
with up to 4 or 8 valves can be installed.
Additionally, the method used to control any
installed pumps can be set.
Note
Depending on the model, up to 2 pumps may
be installed.
For an overview of all available options:
next page.
X-STREAM X2
6.2.3.5, page 6-43)
6.2.3.2.2, page
6-41