Emerson Rosemount Analytical X-STREAM X2 Series Instruction Manual page 220

Gas analyzers
Hide thumbs Also See for Rosemount Analytical X-STREAM X2 Series:
Table of Contents

Advertisement

X-STREAM X2
6.2.3.7
Alarm Setup Menu
Setup..
Alarms..
Level1:
Function:
Level2:
Function:
Page 1
Note!
The unit for concentration values (here: ppm)
is taken from the corresponding line in the
DISPLAY SETUP menu (
6-19).
6-46
6.2 Menu System - Alarm Setup Menu
100 ppm
Low
500 ppm
High
6.2.3.1.3, page
If the system is set up accordingly, the access
code for level 3 must be entered to gain ac-
cess to this menu.
Multi-channel unit:
Select the component to be set in the SELECT
COMPONENT menu.
Sets first concentration alarm level.
Selects alarm output mode for level 1.
Available options:
Off, Low, High, Off FS, Low FS, High FS
(
Chapter 5 includes more details about
these options are alarm settings)
Sets second concentration alarm level.
Selects alarm output mode for level 2.
Available options:
Off, Low, High, Off FS, Low FS, High FS
(
Chapter 5 includes more details about
these options and alarm settings)
Emerson Process Management GmbH & Co. OHG
Instruction Manual
HASX2E-IM-HS
10/2012