X-STREAM X2
Save-Load
CfgData > SvcPort!
SvcPort > CfgData..
Verify!
7-64
7.7. Save / Restore Configuration Data Sets
The SAVE-LOAD.. menu also offers an option
to verify that the saved data on the external
device has not been corrupted during trans-
mission:
With the external device still connected select
the "Verify!" line and press the ENTER key.
The analyzer now loads the data from the ex-
ternal device and compares it to the CfgData.
Finally a status screen shows up, followed by
a COMMAND EXECUTED screen or an error
message, if the data does not match.
Emerson Process Management GmbH & Co. OHG
Instruction Manual
HASX2E-IM-HS
10/2012