7.7.4.3 Restore SvcPort to CfgData
Open the terminal software on your external
device.
Setup..
Save-Load..
Save-Load
CfgData > SvcPort!
SvcPort > CfgData..
Verify!
SvcPort>CfgData
Are you sure?
No!
Yes!
SvcPort > CfgData
-PLEASE WAIT-
for cancel!
8
(i)
-COMMAND EXECUTED-
Save-Load
CfgData > SvcPort!
SvcPort > CfgData..
Verify!
Emerson Process Management GmbH & Co. OHG
7.7. Save / Restore Configuration Data Sets
<-->
Starting at the measurement screen press
the DOWN key to open the MAIN MENU,
enter the SETUP.. and next the SAVE-LOAD..
menu.
If system is setup accordingly, access level 3
code must be entered to gain access to this
menu.
Now select the "SvcPort > CfgData.." line and
press the ENTER key.
A screen appears to confirm the operation:
Select Yes! and after pressing the ENTER
key a new screen shows up showing the
current status.
The instrument now copies the data from the
external device into the analyzer CfgData.
When the procedure has finished, the COM-
MAND EXECUTED screen shows up for 2
seconds, followed by....
... the previous menu page. After a few se-
conds the analyzer automatically starts to
reboot.
X-STREAM X2
7-65