Tab. 3-9: Ir, Uv, Vis, Tcd - Measurement Performance Specifications - Emerson Rosemount Analytical X-STREAM X2 Series Instruction Manual

Gas analyzers
Hide thumbs Also See for Rosemount Analytical X-STREAM X2 Series:
Table of Contents

Advertisement

X-STREAM X2
Measurement Performance Specifications
Detection limit (4 σ)
Linearity
1 4
Zero-point drift
1 4
Span (sensitivity) drift
Repeatability
1 4
Response time (t
)
90
Permissible gas flow
Influence of gas flow
Maximum gas pressure
Influence of pressure
– At constant temperature
– With pressure compensation
Permissible ambient temperature
Influence of temperature
(at constant pressure)
– On zero point
– On span (sensitivity)
Thermostat control
Warm-up time
6
1
Related to full scale
2
Related to measuring value
From gas analyzer inlet at gas flow of 1.0 l/min
3
(electronic damping = 0 s)
4
Constant pressure and temperature
Dependent on integrated photometer bench
5

Tab. 3-9: IR, UV, VIS, TCD - Measurement Performance Specifications

Important Notes Concerning Measurement Specification Data
The specifications given in these tables al-
ways apply to the physical measuring ranges,
as listed e.g. in the INFO-RANGES.. menu
(parameters „MinRange" and „MaxRange").
All performance data is verified during the manufacturing process for each unit by the following
tests:
Linearization and sensitivity test
Long term drift stability test
Climate chamber test
Cross interference test (if applicable)
3-20
3.5 Measurement Specifications
1 4
1 4
3
1 4
≤ 1500 hPa abs. (≤ 7 psig)
8 14
2
7
0 (-20) to +50 °C (32 (-4) to 122 °F)
9
1 13
≤ 5 % (0 to +50 °C / 32 to 122 °F)
6 12
6
Depending on measuring range
7
Pressure sensor is required
8
Limited to atmospheric if internal sample pump
Temperatures below 0 °C (-4 °F) with thermo-
9
stat control only
Thermost. controlled sensor: 75 °C (167 °F)
10
NDIR/UV/VIS
≤ 1 %
≤ 1 %
≤ 2 % per week
≤ 0.5 % per week
≤ 1 %
4 s ≤ t
≤ 7 s
5
90
0.2–1.5 l/min.
≤ 0.5 %
≤ 0.10 % per hPa
≤ 0.01 % per hPa
≤ 1 % per 10 K
none / 60 °C (140 °F)
5
15 to 50 minutes
5
Instruction Manual
Thermal Conductivity (TCD)
≤ 1 %
≤ 1 %
≤ 2 % per week
≤ 1 % per week
≤ 1 %
15 s ≤ t
≤ 30 s
90
0.2–1.5 l/min. ( + 0.1 l/min)
≤ 1 %
11
≤ 1500 hPa abs. (≤ 7 psig)
≤ 0.10 % per hPa
≤ 0.01 % per hPa
0 (-20) to +50 °C (32 (-4) to 122 °F)
≤ 1 % per 10 K
≤ 1 % per 10 K
none / 60 °C (140 °F)
approx. 50 minutes
Note! 1 psi = 68.95 hPa
Flow variation within ± 0.1 l/min
11
Optional thermostatically controlled box with
12
temperature 60 °C (140 °F), not X2GK
13
Temperature variation: ≤ 10 K per hour
14
Special conditions apply to model X2FD
Scaling or zooming (
5-20) cannot improve analog
) cannot improve analog
output specifications to values
better than specified by the phy-
sical measuring ranges
Emerson Process Management GmbH & Co. OHG
HASX2E-IM-HS
10/2012