X-STREAM X2
7.4.1.1.2 External Valve to Digital Output Assignment
Setup..
In/Outputs..
DigitalOutputs..
Output1:
Output2:
Output3:
Output4:
Page 1
Output5:
Output6:
Output7:
Output8:
Page 2
7-14
7.4.1 Preparing Calibrations
Failure
MaintRequ
OffSpec
FctCheck
Sample
V1
V2
V3
If your analyzer has to control external valves,
at first check if all valves required for calibra-
tion are connected to digital outputs.
Then open the SETUP - IN/OUTPUTS -
DIGITALOUTPUTS menu to software assign
the valves to the outputs:
This menu configures the digital outputs: All
outputs (default and optional) support the
same range of signals/functions. "Outputs
1-4" are available in every unit and by default
setup to provide NAMUR signals (see figures
to the left).
Further pages are indicated by a down arrow
(
) only when at least one extension card
(outputs 5 - 13) is installed
"Outputs 5 - 13" are present on the first ex-
tension card (outputs 9 to 13 on menu pages
3 & 4 not shown in this example).
Note!
Depending on the analyzer model, 1 or 2
Digital I/O extension cards can be installed.
The next step is to assign the valves to the
channels: Continue with
7-13.