Instruction Manual
HASX2E-IM-HS
10/2012
6.2.2.2
Span calibration menu
Control..
Span calibration..
Cancel calibration!
Start calibration!
Span gas
2000.0 ppm
CO2.1
Page 1
RESET..
Status..
CO2.1
Page 2
Note!
For further details on calibration procedures,
see
chapter 7 Maintenance.
Emerson Process Management GmbH & Co. OHG
6.2 Menu System - Control menu
134.1 ppm
134.1 ppm
Multi-channel unit:
Select the channel to be calibrated in the
SELECT COMPONENT menu.
Pressing the 8 key in this line will abort the
span calibration and cancel any changes.
Pressing the 8 key in this line will start the
span calibration.
Current nominal span gas concentration
(span gas concentration can be set in the
SETUP menu
6.2.3.2.1, page 6-24)
Current measured gas concentration.
Press the 8 key in this line to enter a subme-
nu to reset the calibration parameters (
6.2.2.2.1, page 6-8)
Press the 8 key in this line to view the current
calibration status (
Current measured gas concentration.
Multi-channel unit:
Press the  key to enter the SELECT COM-
PONENT menu to change the settings for a
different channel.
X-STREAM X2
6.2.2.2.2, page 6-9)
6-7