Interfaces, signal inputs and outputs
Interface signals are accessed in different ways depending on the analyzer model. (This does
not apply to Ethernet, which always uses RJ45):
X-STREAM X2GK, X2GP: standard:
X-STREAM X2XF, X2FD:
All models are supplied with
up to 4 analog outputs
(Standard: 1 analog output per channel)
4 relay outputs
Each output can be configured to one of the
Emerson Process Management GmbH & Co. OHG
2.1 Common Technical Data
optional:
following functions:
X-STREAM X2
subminD plugs and sockets
screw-type terminal adapters
internal screw-type terminals
4 (0)–20 mA (R
≤ 500 Ω)
B
electrically isolated from each other and
from other electronic components;
user-configurable activation and
deactivation concentration levels;
support for NAMUR NE 43 operation
modes, configurable via keypad and
Modbus
dry contacts,
max. load. 30 V; 1 A; 30 W resistive
NAMUR NE 107 status signal
"Failure"
"Maintenance request"
"Out of specification"
"Function check"
1 of 2 concentration limits per channel,
Control signals for
external valve V1...V8,
external sample gas valve
external pump
Zoom status for analog outputs
2-3