4.6.3 Installation - X-STREAM X2XF Field Housings
Signal inputs and outputs
The number of actually available signal out-
puts, and also the number of built-in modules
with screw-type terminals, varies according to
the configuration of the unit.
Characteristics of terminals:
Recommended wire gauge:
Skinning length:
Thread:
Min. tightening torque:
Verify the power supply at installation site meets the specification given on
the analyzer´s nameplate label, before installing the instrument!
Verify power cables are disconnected and/or instrument is de-energized
prior to working at the terminals!
Verify the power cord is layed with a distance of at least 1 cm (0.4 in) to any
signal cable to ensure proper insulation from signal circuits!
Emerson Process Management GmbH & Co. OHG
A maximum of three modules with 36 termi-
nals each can be fitted.
The terminals can be accessed by opening
the front panel of the instrument.
ELECTRICAL SHOCK HAZARD
X-STREAM X2
0.14...1.5 mm
(AWG 26...AWG 16),
2
end sleeves not required
5 mm (0.2")
M2
0.25 Nm (2.21 in.lb)
4-27