X-STREAM X2
P4
8-40
8.4
Troubleshooting on Components
LB3
R119
Step 5:
To adjust the physical span:
Apply span gas to the analyzer.
Do not disconnect the DVM:
P4.5 Raw signal (+)
P4.15 Bridge voltage (-); GND
CAUTION!
Do not short-circuit pins!
P4.15
P4.5 Raw
Adjust the voltage to 10V
utilizing R119.
If 10V is not within the adjustable
range, it is necessary to change
the signal amplification with sol-
der bridge LB3:
For an amplification close
factor of
20
150
300
500
Step 6:
Now once more check the zero
point:
Apply zero gas to the analyzer.
Do not disconnect the DVM:
The voltage should be 0 V ±
500 mV.
If it does not, repeat from step 3!
Emerson Process Management GmbH & Co. OHG
Instruction Manual
HASX2E-IM-HS
10/2012
1-5
3-5
4-5
2-3-4-5