Fig. 2-13: X-Stream X2Fd - Power Supply Terminals / Fuse Holders - Emerson Rosemount Analytical X-STREAM X2 Series Instruction Manual

Gas analyzers
Hide thumbs Also See for Rosemount Analytical X-STREAM X2 Series:
Table of Contents

Advertisement

X-STREAM X2
2.2.4 Model-Specific Technical Data: X-STREAM X2FD
Power supply
Nominal input voltage
Input voltage range
Nominal input current
standard
with temperature control
Cable cross-section:
Cable entry via
Permissible cable outer diameter:
Power supply fuses
Fuse holders are integrated within power
supply terminals.
Fuse ratings:
1
Power terminals with integrated
fuse holders
2
Protective earth terminal (PE)
3
Power cable entry
4
EMI power supply filter
Fig. 2-13:
X-STREAM X2FD - Power Supply Terminals / Fuse Holders
2-20
100–240 V
50/60 Hz,
global power supply unit
85–264 V , 47–63 Hz
1.3–0.7 A max.
3–1.5 A max.
Connection via touch proof screw-type
terminals located inside the unit near
cable glands (Fig. 2-13).
max. 4 mm
(10 AWG), end sleeves not
2
required.
1 cable gland, classified IP 68 or suitable
conduit with metric-to-NPT adaptor
3–13 mm (0.11–0.5 inch),
dependent on inset used in cable gland
AC 230 V / T 4 A / 5x20 mm
4
2
Emerson Process Management GmbH & Co. OHG
Instruction Manual
HASX2E-IM-HS
10/2012